Masimo® Plethysmograph Variability Index as a Tool for Assessment of Fluid Responsiveness in Elective Major Abdominal Surgeries

Essam, Ahmed and Ali, Mohamed and Maher, Mohamed and Mokhtar, Ali and Omar, Sohila and El-Sabae, Hossam and Hafez, Mohamed (2016) Masimo® Plethysmograph Variability Index as a Tool for Assessment of Fluid Responsiveness in Elective Major Abdominal Surgeries. British Journal of Medicine and Medical Research, 16 (2). pp. 1-11. ISSN 22310614

[thumbnail of Ali1622016BJMMR26030.pdf] Text
Ali1622016BJMMR26030.pdf - Published Version

Download (371kB)

Abstract

Background: Maximizing the stroke volume (SV) as measured by Trans-oesophageal Doppler (TED) optimizes preload, & is a goal-directed fluid therapy technique that has been used in a variety of clinical settings. Masimo® Plethysmograph variability Index (PVI) is a reliable, safe & noninvasive tool to guide fluid management. PVI is an automated measure of the dynamic change in the perfusion index (PI) that occurs during a respiratory cycle.

This study was designed to determine whether PVI, measured using finger co-oximetry is an efficient predictor of fluid responsiveness in low-risk patients undergoing elective major abdominal surgery.

Subjects and Methods: 60 ASA I-II patients of either sex, 25-60 years old, undergoing major abdominal surgery were enrolled in this study. A Masimo® Radical-7 Pulse Co-Oximeter probe & a Cardio Q TED probe were applied to each patient. In all patients, a fluid bolus of 500 ml of 130/0.4 tetrastarch colloid solution was administered rapidly via pressurized IV infusion. Maintenance & deficits were calculated routinely. If the SV decreased by 10%, a 250-mL bolus of colloid was given via fast infusion. Patients’ demography, TED-derived measurements: (SV & Flow Time corrected (FTc)), Masimo®-derived measurements: (PVI & PI), HR and MAP were all collected and statistically analyzed. Measurements were done at five minutes post-induction T1, Ten minutes after volume expansion (500 ml colloid) T2, If the SV decreased by 10%, (guided by TED) T3, Then 250 ml colloid is given. Ten minutes after a 250-ml colloid bolus T4.

Results: A significant difference was found in FTc, SV, PI & PVI in T1 vs. T2 & T3 vs. T4 (P=0.001). There was a significant difference in PI & PVI between responders & non-responders for the 1st bolus (P<0.05) and in SV & PVI in subsequent boluses (P<0.01). There was no significant difference between percent changes of SV and PVI at T3 & T4.

Conclusions: Plethysmograph Variability Index (PVI) measured by Masimo® Co-Oximeter is an efficient predictor of fluid responsiveness as SV measured by TED in low risk patients undergoing elective major surgery.

Item Type: Article
Subjects: STM One > Medical Science
Depositing User: Unnamed user with email support@stmone.org
Date Deposited: 17 May 2023 11:07
Last Modified: 04 Sep 2024 04:04
URI: http://publications.openuniversitystm.com/id/eprint/1130

Actions (login required)

View Item
View Item